Can transcutaneous oxygen and carbon dioxide tensions reflect arterial levels during exercise testing?

This lab study shows that transcutaneous values can closely follow changes in arterial blood gas by using an electrode temperature of 45 °C to monitor exercise with gradual workload.

INTRODUCTION

The measurement of arterial oxygen tension (pO2(a)) and carbon dioxide tension (pCO2(a)) are important and widely used tests in clinical patient monitoring and in the pulmonary function laboratory for assessing patients before and after exercise testing. 

Arterial blood is obtained by puncture or from an indwelling cannula, but the measurements of pO2(a) and pCO2(a) in blood sampled intermittently have the serious disadvantage that critical changes may be missed, especially if a patient is assessed with an arterial sample obtained immediately post-exercise where the blood sampled may not reflect pulmonary gas exchange during exercise [1]. 

Arterial punctures are therefore impractical for continuous monitoring, while indwelling arterial cannulas may be uncomfortable and are potentially hazardous [2]. 

It would be advantageous, therefore, to have available a system which could monitor gas exchange in exercise without the potential morbidity associated with arterial cannulation.

The development of a combined transcutaneous sensor for the non-invasive, continuous measurement of oxygen tension (tcpO2) and carbon dioxide tensions (tcpCO2) at the skin surface has stimulated interest in the information that could be derived from the simultaneous monitoring of pO2(a) and pCO2(a) in the adult patient during exercise testing.

Several previous studies have demonstrated that the transcutaneous oxygen electrode provides a reliable method of continuously monitoring the changes in arterial oxygen tensions of adults undergoing exercise testing [3,4,5], but differences were found in other studies [6]. 

Some of these differences can be accounted for by the different temperature settings used with the transcutaneous electrode (42-44 oC) producing a range of response times, or whether an in vivo calibration procedure was performed using a single arterial stab or earlobe capillary measurement prior to the exercise test.

More controversy surrounds the use of the transcutaneous carbon dioxide monitor in the monitoring of arterial carbon dioxide tensions during exercise testing. It has been shown that tcpCO2 accurately reflects arterial 

pCO2(a) in hemodynamically stable patients at rest using an electrode temperature of 39-40 oC. At this temperature, however, there was a slow response time for the monitoring of changes which, in some cases, was up to three minutes, which would preclude its use in exercise testing [7]. 

Higher temperatures of the electrodes produce an increase in the temperature of capillary blood which leads to a systematically higher skin surface carbon dioxide ten­­si­on than would be the case if measured at 37 oC. 

It is now possible to incorporate a temperature correction factor based on the Siggaard-Andersen equation [8] to transcutaneous monitors, which allows the tcpCO2 electrode to be used at a higher temperature with the tcpCO2 reading corrected to 37 oC. 

The use of the higher temperature has the advantage that it improves the response time of the tcpCO2 electrode, so that more rapid changes in arterial carbon dioxide tensions can be followed.

TRANSCUTANEOUS MONITORING IN THE ADULT PATIENT: A PERSONAL EXPERIENCE

It has previously been shown, in our own laboratory, that the oxygen and carbon dioxide tension measured by the transcutaneous method correlates with the direct measurement of pO2(a) and pCO2(a) in hemodynamically stable patients at rest using an electrode temperature of 45 oC [9]. 

This confirms the findings of previous studies [10]. In 42 subjects, who were referred from a respiratory clinic for arterial blood sampling, the pCO2(a) ranged from 3.07 to 7.47 kPa. These values were compared to those obtained from a transcutaneous monitor following an in vitro calibration procedure using standard gas mixtures. 

The site of attachment chosen for the electrode was the flexor surface of the forearm, which has a high capillary density and a relatively thin epidermis and is therefore a convenient site for measuring transcutaneous gases. 

There was a significant correlation between tcpCO2(37 oC) and pCO2(a) derived from these 42 simultaneous measurements (r=0.98; p<0.001). The mean tcpCO2 and pCO2(a) were 5.0 +/- 0.15 kPa and 5.05 +/- 0.15 kPa, respectively. 

An analysis of Bland and Altmann showed a mean difference of 0.02 kPa  with no significant difference between the two measurements.

In the same group of patients the pO2(a) ranged from 6.8 to 14.0 kPa. Although there was a significant correlation between the two measurements (r=0.91; p<0.001), the transcutaneous level of pO2 was generally lower than the simultaneous measurement of pO2(a). 

There were, however, large inter-individual differences between the two measurements. The analysis of Bland and Altmann showed a mean difference of 0.89 kPa. The mean tcpO2 and pO2(a) were 9.3 +/- 0.25 kPa and 10.4 +/- 0.28 kPa, respectively. 

Analysis of variance showed a significant difference between these two methods of measurement (F=7.9; p=0.006). This difference is due to a number of factors as described by Hutchinson et al [11].

The oxygen pressure recorded at the transcutaneous electrode will tend to be lower than the true arterial pO2 because of oxygen consumption by the skin itself. Conversely, the temperature of the skin underlying the electrode is several degrees above that of normal blood. 

This displaces the oxyhemoglobin dissociation curve to the right, producing an increase in the pO2(a) at a given hemoglobin saturation. Thus, the final tcpO2 reflects a balance between these two factors. 

While a good correlation between tcpO2 and pO2(a) is evident from these results in adults, the scatter of results shows that the tcpO2 cannot predict the pO2(a) with consistent accuracy when in vitro methods are used to calibrate the transcutaneous monitor. 

The performance of an in vivo calibration using a sample of arterial or arterialized earlobe capillary blood has been shown to improve the accuracy of the tcpO2 monitor by compensating for the difference between pO2(a) in individual patients [12].

THE APPLICATION OF TRANSCUTANEOUS MONITORING OF pO2 DURING CARDIOPULMONARY EXERCISE TESTING

The temperature of 45 oC used produces response times of the tcpO2 and tcpCO2 electrode to a change in breathing pattern or inspired gas of 30 and 55 seconds, respectively. 

These response times are of an order which has allowed the investigation of the use of the combined electrode to non-invasively monitor arterial blood gasses during exercise testing. We have therefore validated the transcutaneous monitor in 24 patients referred for progressive cardiopulmonary exercise testing against direct arterial sampling from an indwelling arterial cannula [13]. 

The group comprised patients with a range of cardiopulmonary disorders of varying degrees of severity, and also included patients with unexplained breathlessness. The transcutaneous electrode was firstly calibrated using standard gas mixtures. 

The electrode was then attached to the skin and when a steady tcpO2 and tcpCO2 had been achieved (after about ten minutes), an arterial blood sample was obtained from a previously inserted arterial cannula, and pO2 and pCO2 measured. The gain controls of the device were then altered so that the output of the tcpO2 and tcpCO2 corresponded to the measured pO2(a) and pCO2(a) (in vivo calibration procedure).

A progressive exercise test was then perform­ed using an electrically braked bicycle ergometer. The tcpO2 and tcpCO2 were monitored for two minutes while the patient was seated at rest. 

An arterial blood sample was obtained at the midpoint of this period, and, taking into account the response time of the transcutaneous electrode, the transcutaneous monitor readings were recorded at the end of this period. 

The patient was then asked to exercise for as long as possible until symptomatic limitation. During the first two minutes of exercise no additional load was applied, and an arterial blood sample was obtained at the midpoint of this period.

Transcutaneous values were recorded at the end of this period. Thereafter, the work load was increased by 10-25 watts, depending on the individual patient, every two minutes until symptomatic limitation. 

Blood samples were taken at the midpoint and transcutaneous values taken at the end of each work load period. In these 24 patients 140 simultaneous measurements of arterial blood gases and transcutaneous values were obtained following the in vivo calibration. The range of arterial pO2 measured was 6.1 to 14.2 kPa. 

The analysis of Bland and Altmann on the two pO2 measurements showed a mean difference of 0.08 kPa. The mean tcpO2 and pO2(a) were 11.5 +/- 0.35 kPa and 11.5 +/- 0.33 kPa, respectively. Analysis of variance shows that there is no significant difference between these measurements (F=0.05; p=0.94). The in vivo calibration procedure produces an even scatter of results about the mean value with good limits of agreement suggesting that there is no difference between the measurement of tcpO2 and pO2(a) on exercise testing.

The in vivo calibration procedure produced no adverse affects on the good relationship between tcpCO2 and pCO2(a) already described. The pCO2(a) ranged from 3.6 to 7.6 kPa. 

The analysis of Bland and Altmann showed a mean difference of 0.02 kPa. The mean tcpCO2 and pCO2(a) were 5.6 +/- 0.2 and 5.6 +/- 0.2 kPa, respectively, with no significant difference between the two measurements (F=0.06; p=0.89).

In all the exercise tests the trend of gas exchange as measured by the transcutaneous monitor was true to the trend as measured from direct arterial sampling. It might be suggested that the necessity of obtaining an arterial blood sample of arterialized earlobe capillary sample to perform the initial in vivo calibration would invalidate the non-invasive principle of the transcutaneous monitor. 

However, a single arterial puncture or earlobe capillary sample carefully performed is a procedure with a very low complication rate, and once this has been completed, the usefulness of the transcutaneous monitor can be demonstrated.

CONCLUSION

The performance of an in vivo calibration procedure produces values of transcutaneous pO2 and pCO2 which are not significantly different from arterial blood gases during an exercise test. 

It has been suggested that the slow response characteristics of transcutaneous monitors would render the use of a transcutaneous monitor unsuitable for exercise testing. 

We have shown in our laboratory, however, that using the highest recommended temperature of 45 oC, producing a response time of 55 seconds for the CO2 electrode, and combining this with an exercise protocol of gradual work load increments of two-minute periods to prevent abrupt and large changes in blood gas tensions allows changes in arterial blood gases to be closely followed by transcutaneous values.

The Respiratory Physiology Laboratory at Glasgow Royal Infirmary now routinely uses transcutaneous monitoring, after an in vivo calibration procedure using a single arterial stab or arterialized earlobe capillary sampling, to monitor changes in arterial blood gas values non-invasively during exercise testing. 

When this is combined with mixed expired gas analysis this allows the calculation of sensitive indices of gas exchange including the alveolar-arterial oxygen gradient and the dead space/tidal volume ratio to be monitored throughout the exercise test. 

The transcutaneous monitor has been tolerated without any associated morbidity in a wide range of patients who have been exercised in the laboratory [13,14,15,16].

References
  1. Ries AL, Fedullo PF, Clausen JL. Rapid changes in arterial blood gas levels after exercise in pulmonary patients. Chest 1983; 83: 454-56.
  2. Bedford RF, Wollman WH. Complications of percutaneous radial artery cannulation: an objective prospective study in man. Anaesthesiology 1973; 38: 228-36.
  3. McDowell JW, Thiede WH. Usefulness of the transcutaneous pO2 monitor during exercise testing in adults. Chest 1980; 78: 853-55.
  4. Schonfield T, Sargent CW, Bautista D, et al.  Transcutaneous oxygen monitoring during exercise testing. Am Rev Respir Dis 1980; 121: 457-62.
  5. Hughes JA, Gray BJ, Hutchison DCS. Changes in transcutaneous oxygen tension during exercise in pulmonary emphysema. Thorax 1984; 39: 424-31.
  6. Steinacker JM, Wodick RE. Transcutaneous pO2 during exercise. Adv Exp Med Biol 1984; 169: 763-74.
  7. Nickerson BG, Paterson C, McCrea R, Monaco F. In vivo response times for a heated skin surface CO2 electrode during rest and exercise. Ped Pulmonol 1986; 2: 135-40.
  8. Siggaard-Andersen O. The Acid-Base Status of Blood. Copenhagen: Munksgaard, 1976: 89.
  9. Carter R. The measurement of transcutaneous oxygen and carbon dioxide tensions in the adult patient. Care Crit III 1988; 4: 13-15.
  10. Mahutte CK, Michiels TM, Massell KT. Trueblood DM. Evaluation of a single transcutaneous pO2 - pCO2  sensor in adult patients. Crit Care Med 1984; 12: 1063-66.
  11. Hutchison DCS, Rocca G, Honeybourne D. Estimation of arterial oxygen tension in adult subjects using a transcutaneous electrode. Thorax 1981; 36: 473-77.
  12. Gray BJ, Heaton RW, Henderson A, Hutchison DCS. In vivo calibration of a transcutaneous oxygen electrode in adult patients. In: Huch A, Huch R, Rooth G, eds. Continuous Transcutaneous Monitoring. Advances in Experimental Medicine and Biology. New York and London: Plenum Press, 1987: 75-78.
  13. Sridhar MK, Carter R, Moran F, Banham SW. Use of a combined oxygen and carbon dioxide transcutaneous electrode in the estimation of gas exchange during exercise. Thorax 1993; 48: 643-47.
  14. Carter R, Naik S, Stevenson RD, Richness D, Wheatley DJ. Breathlessness and ventilatory abnormalities in patients with cardiac failure awaiting transplantation. J Heart and Lung Transplantation 1994; 13: 86. Abstr. 217.
  15. Tweddel A, Carter R, Banham  SW, Hutton I. Breathlessness in Microvascular Angina. Resp Med 1994; 88: 731-36.
  16. Sridhar MK, Carter R, Banham SW, Moran F. An evaluation of integrated cardiopulmonary exercise testing in a pulmonary function laboratory. Scot Med J 1995;  40: 113-116.
+ View more
References
  1. Ries AL, Fedullo PF, Clausen JL. Rapid changes in arterial blood gas levels after exercise in pulmonary patients. Chest 1983; 83: 454-56.
  2. Bedford RF, Wollman WH. Complications of percutaneous radial artery cannulation: an objective prospective study in man. Anaesthesiology 1973; 38: 228-36.
  3. McDowell JW, Thiede WH. Usefulness of the transcutaneous pO2 monitor during exercise testing in adults. Chest 1980; 78: 853-55.
  4. Schonfield T, Sargent CW, Bautista D, et al.  Transcutaneous oxygen monitoring during exercise testing. Am Rev Respir Dis 1980; 121: 457-62.
  5. Hughes JA, Gray BJ, Hutchison DCS. Changes in transcutaneous oxygen tension during exercise in pulmonary emphysema. Thorax 1984; 39: 424-31.
  6. Steinacker JM, Wodick RE. Transcutaneous pO2 during exercise. Adv Exp Med Biol 1984; 169: 763-74.
  7. Nickerson BG, Paterson C, McCrea R, Monaco F. In vivo response times for a heated skin surface CO2 electrode during rest and exercise. Ped Pulmonol 1986; 2: 135-40.
  8. Siggaard-Andersen O. The Acid-Base Status of Blood. Copenhagen: Munksgaard, 1976: 89.
  9. Carter R. The measurement of transcutaneous oxygen and carbon dioxide tensions in the adult patient. Care Crit III 1988; 4: 13-15.
  10. Mahutte CK, Michiels TM, Massell KT. Trueblood DM. Evaluation of a single transcutaneous pO2 - pCO2  sensor in adult patients. Crit Care Med 1984; 12: 1063-66.
  11. Hutchison DCS, Rocca G, Honeybourne D. Estimation of arterial oxygen tension in adult subjects using a transcutaneous electrode. Thorax 1981; 36: 473-77.
  12. Gray BJ, Heaton RW, Henderson A, Hutchison DCS. In vivo calibration of a transcutaneous oxygen electrode in adult patients. In: Huch A, Huch R, Rooth G, eds. Continuous Transcutaneous Monitoring. Advances in Experimental Medicine and Biology. New York and London: Plenum Press, 1987: 75-78.
  13. Sridhar MK, Carter R, Moran F, Banham SW. Use of a combined oxygen and carbon dioxide transcutaneous electrode in the estimation of gas exchange during exercise. Thorax 1993; 48: 643-47.
  14. Carter R, Naik S, Stevenson RD, Richness D, Wheatley DJ. Breathlessness and ventilatory abnormalities in patients with cardiac failure awaiting transplantation. J Heart and Lung Transplantation 1994; 13: 86. Abstr. 217.
  15. Tweddel A, Carter R, Banham  SW, Hutton I. Breathlessness in Microvascular Angina. Resp Med 1994; 88: 731-36.
  16. Sridhar MK, Carter R, Banham SW, Moran F. An evaluation of integrated cardiopulmonary exercise testing in a pulmonary function laboratory. Scot Med J 1995;  40: 113-116.
Disclaimer

May contain information that is not supported by performance and intended use claims of Radiometer's products. See also Legal info.

No portrait of author Roger Carter

 

Department of respiratory medicine 
Glasgow Royal Infirmary 
Glasgow 
Scotland

Articles by this author
Acutecaretesting handbook

Acute care testing handbook

Get the acute care testing handbook

Your practical guide to critical parameters in acute care testing. 

Download now
Webinar

Related webinar

Role of transcutaneous CO2 monitoring in high risk respiratory patients

Presented by Gil De Oliveira, MD

Watch the webinar
Webinar on demand: Evolution of blood gas testing - Part 1

Related webinar

Evolution of blood gas testing Part 1

Presented by Ellis Jacobs, PhD, Assoc. Professor of Pathology, NYU School of Medicine.

Watch the webinar
Webinar: Evolution of blood gas testing - Part 2

Related webinar

Evolution of blood gas testing Part 2

Presented by Ellis Jacobs, PhD, Assoc. Professor of Pathology, NYU School of Medicine.

Watch the webinar
Preanalytical errors handbook

Blood gas Preanalytics app

Get the Blood gas Preanalytics app for your smartphone

This smartphone app focuses on the preanalytical phase of blood gas testing and what operators can do to avoid errors.

Download app

Sign up for the Acute Care Testing newsletter

Sign up
About this site About Radiometer Contact us Legal notice Privacy Policy
This site uses cookies Read more