Printed from acutecaretesting.org
June 2002
Fetal capillary blood pH (fetal blood sampling)
During labor the fetus is extremely at risk as compromised oxygen supply from the placental circulation to the fetus may develop either gradually or suddenly. Such a situation can result in severe organ damage, especially cerebral damage, or death of the fetus if not treated adequately.
It is therefore of major importance to monitor the fetus during labor to detect any sign of compromised fetal health demanding therapeutic intervention. During the last century the fetal heart rate has been used as an indicator of intrauterine well-being.
A continuous simultaneous recording of fetal heart rate and uterine activity (CardioTocoGraphy, CTG) is based on studies performed by Hon (1958) [1], and this method eliminates the human error in counting the heart beats, and both short-term and long-term variability are more strictly evaluated.
Nevertheless, these methods have proven an insufficient sensitivity and specificity in diagnosing fetal asphyxia.
In the 1970s and 1980s several randomized studies concerning fetal monitoring with cardiotocography as the only monitoring equipment were performed, and in most cases the results showed only a very modest decrease in morbidity of the newborn, while the rate of instrumental delivery (especially Cesarean section) increased.
Therefore, it is obvious that another or a supplementary fetal monitoring is necessary to diagnose fetal asphyxia.
FETAL ASPHYXIA
The term asphyxia is not exactly defined in the literature, although the condition is well known by obstetricians and neonatologists. To most obstetricians the condition means a state of the fetus with depressed vital functions, usually defined by a fetal acidosis combined with hypoxia/anoxia.
Signs of asphyxia during labor are abnormal heart rate, pathological CTG readings, green meconium-stained amniotic fluid, and low scalp-blood pH.
ACID-BASE CHANGES DURING FETAL HYPOXIA
During the first stage of fetal hypoxia, caused by partial compression of the umbilical cord, by a decreased fetal cardiac output, or by placental insufficiency, the most evident change is an increasing pCO2.
The hypoxia is at this stage not compromising the aerobic metabolism of carbohydrate to CO2 and H2O. An acid-base evaluation of the blood will show a respiratory acidosis with normal base excess value.
During the second and more dangerous stage of fetal hypoxia, the oxygen delivery to the fetus becomes insufficient, and the metabolism of carbohydrate changes to an anaerobic metabolism with the production of lactic acid.
A distinct decrease of pH occurs during oxygen deficiency, until the low pH itself stops all further metabolic activities. Consequently, the acidosis will only be progressive for a short time, the slope of the pH decrease being an indicator of the oxygen deficiency.
Usually the fetal asphyxia develops during the described two-step mechanism. In some cases, the first stage is prolonged, and it may eventually be interrupted, if normal oxygen exchange between the fetus and the mother is reestablished.
However, in other cases, with a sudden total occlusion of the umbilical cord or a total placental abruption, the hypoxia may be so evident that the anaerobic metabolism starts immediately after the obstetric complication.
NORMAL VALUES OF FETAL pH
Initially during the first stage of labor the pH decreases slowly (0.016 pH units per hour) while during the second stage of an even normal labor the pH value decreases faster (0.12 pH unit per hour) [2]. Therefore, normal fetal capillary-blood pH ranges from 7.25 to 7.45 during labor.
FETAL BLOOD SAMPLING
The method of fetal blood sampling was described by Saling in 1961 [3]:
- The membranes must be ruptured and the cervical os dilated more than 3-4 cm.
- The presenting part of the fetus is visualized by an amnioscope. The amnioscope should be pressed gently against the skin to avoid leak of amniotic fluid into the area but still permitting normal vascularization of the skin.
- The skin is cleaned and dried with swaps. Rubbing several times will assure increased blood flow to the area. To avoid the blood drops from spreading, silicone is placed on the swap and spread over the area.
- An incision is made in the upper part of the area with a 2-mm blade. In some cases 2 or 3 incisions are necessary to produce adequate bleeding.
- The blood drops are collected into a preheparinized capillary tube. The capillary tube tip should be placed directly in the drop and the tube tilted slightly with the external end lower than the internal to facilitate the filling of the tube. Suction on the tube should be avoided.
- The blood is transferred to the pH meter and the value measured.
The volume of blood obtained performing fetal scalp-blood sampling is usually limited and in many cases not large enough to perform a total acid-base evaluation including pH, pCO2 and base excess. Measurement of blood pH alone can be performed in small blood samples, and in clinical practice this monitoring procedure is preferred during labor. Furthermore, a pH decrease will always initiate the decrease in base excess, and this early warning is of importance for the obstetrician.
CLINICAL APPLICATION OF INTRAPARTUM FETAL CAPILLARY pH
Fetal capillary-blood pH values above 7.25 indicate that the fetus at the time of the sampling does not suffer from hypoxia. Labor should be allowed to progress, but if the heart rate or the cardiotocogram indicates pathological changes, the blood sampling should be repeated.
pH values between 7.20 and 7.25 are signs of prepathological condition. Measures should be taken to improve fetal oxygenation (change of maternal position, oxygen mask). If the fetus is not delivered 15 min after the pH measurement was performed, a subsequent fetal blood sampling must be performed. If the new pH value is still between 7.20 and 7.25, new blood sampling should be done every 15-30 min.
A capillary-blood pH value below 7.20 indicates a pathological fetal acidosis, and the fetus should be delivered immediately.
BENEFITS OF pH MEASUREMENT DURING LABOR
Continuous electronic monitoring with the cardiotocograph alone and electronic monitoring supplemented with fetal blood sampling and pH measurements have been evaluated in randomized studies.
The rate of instrumental deliveries (Cesarean sections, forceps and vacuum extractions) was lower when fetal blood sampling was performed, and the number of newborns with asphyxia was equal in both groups [4].
The reduction of Cesarean-section rate causes a decrease in the maternal complications following delivery. From an economical point of view, the costs of introducing pH measurements on fetal blood sampling are low in relation to the savings from a lower Cesarean-section rate [5].
COMPLICATIONS FROM FETAL BLOOD SAMPLING
Very few complications following fetal blood sampling have been reported. Prolonged bleeding after blood sampling may be caused by a fetal coagulation disorder or an extensive laceration.
Lacerations were described previously, but with the use of new smaller knifes, these complications are rare. The incision site should, however, be inspected during the following two uterine contractions to ensure that the bleeding has stopped.
In the case of bleeding, compression with a swab will usually stop the bleeding. Infection in the incision wound may happen, but all complications are seldom, in most materials much lower than 1 per cent.
TROUBLE SHOOTING |
|||
PROBLEM |
POSSIBLE CAUSE |
SOLUTION |
|
1 |
Not sufficient
|
Too small incision |
Larger incision |
Insufficient blood
|
Rubbing the skin
|
||
2 |
Clotting of the
|
Too slow filling of
|
See point 1 |
Tissue tromboplastin
|
Use a sharp pointed
|
||
Insufficient heparinization |
Use correct preheparinized
|
||
If not immediately
|
|||
3 |
Air bubbles in the blood sample |
Incorrect sampling procedure |
Do not remove the
|
References+ View more
- Hon EH. The electronic evaluation of the fetal heart. Am J Obstet Gynecol 1958; 75: 1215-30.
- Weber T, Hahn-Pedersen S. Normal values for fetal scalp tissue pH during labour. Br J Obstet Gynaecol 1979; 86: 728-31.
- Saling E. Das Kind im Bereich der Geburtshilfe. Stuttgart: Georg Thieme Verlag, 1966.
- Haverkamp AD, Orleans M, Langendoerfer S, McFee J, Murphy J, Thompson HE. A controlled trial of the differential effects of intrapartum fetal monitoring. Am J Obstet Gynecol 1979; 134: 399-408.
- Mugford M. The cost of continuous electronic fetal monitoring in low risk labour. In: Spencer JAD, Ward RHT, eds. Intrapartum fetal surveillance. London: RCOG press, 1993: 241-52.
References
- Hon EH. The electronic evaluation of the fetal heart. Am J Obstet Gynecol 1958; 75: 1215-30.
- Weber T, Hahn-Pedersen S. Normal values for fetal scalp tissue pH during labour. Br J Obstet Gynaecol 1979; 86: 728-31.
- Saling E. Das Kind im Bereich der Geburtshilfe. Stuttgart: Georg Thieme Verlag, 1966.
- Haverkamp AD, Orleans M, Langendoerfer S, McFee J, Murphy J, Thompson HE. A controlled trial of the differential effects of intrapartum fetal monitoring. Am J Obstet Gynecol 1979; 134: 399-408.
- Mugford M. The cost of continuous electronic fetal monitoring in low risk labour. In: Spencer JAD, Ward RHT, eds. Intrapartum fetal surveillance. London: RCOG press, 1993: 241-52.
May contain information that is not supported by performance and intended use claims of Radiometer's products. See also Legal info.
Acute care testing handbook
Get the acute care testing handbook
Your practical guide to critical parameters in acute care testing.
Download nowRelated webinar
Evolution of blood gas testing Part 1
Presented by Ellis Jacobs, PhD, Assoc. Professor of Pathology, NYU School of Medicine.
Watch the webinar